| Vessel position | 6 standard (7, 8 optional) |
|---|
| Medium volume | 500–1,000 mL; 100, 200 and 250 mL with small-volume kits |
|---|
| Methods | USP 1 (basket), 2 (paddle), 5 (paddle over disk), 6 (rotating cylinder); Intrinsic, immersion cell, felodipine baskets, suppository baskets |
|---|
| Speed range (rpm) | 25–300 (25–30 for qualification only); optional 2 speeds |
|---|
| Speed control | Digital closed loop; typically ±2 rpm (up to 100), ±1% above 100 rpm |
|---|
| Temperature range (°C) | Ambient (max. 28) – 50; optional independent cooling down to 5 |
|---|
| Temperature stability (°C) | ±0.1 |
|---|
| Temperature sensor | PT100, DIN 751 class A (resolution 0.01 °C, accuracy ±0.2 °C); optional additional in-vessel sensor |
|---|
| Display | 7 inch colour touchscreen (usable with gloves) |
|---|
| Interface | USB, LAN; browser interface |
|---|
| Dimensions (W × D × H) (mm) | 660 × 430 × 660 (heater: 260 × 130 × 370) |
|---|
| Weight (without accessories) (kg) | 37 (water tank empty); 52 (full) |
|---|
| Supply | Main unit 220–240 V, 50/60 Hz, 0.10 kW; heater 230 V or 115 V 1.85 kW, 100 V 1.50 kW |
|---|