Instruments, consumables and technical service for the pharmaceutical industry and laboratories

+90 216 514 50 36

/ Mon–Fri 09:00–18:00

Microtube rotors
Product / Centrifuges

Designed for the centrifugation of smaller sample sizes, from 0.5 ml to 2 ml tubes and PCR strips. Programmable in 10 RCF increments up to 22,000 RCF, with a maximum radius of 8.75 cm and a 45° tube angle. A sealed lid is included; RS04 and RS05 reducers fit some models.

Manufacturer website (opens in a new tab)

Source: manufacturer website (opens in a new tab) · Image: manufacturer website

Order codes and options

4 codes

Microtube rotors order codes and properties
CodeRotor typeLargest tubeRCF rangeMax. speedCompatible seriesReducerNoteQuote list
C-R-FA542424 × 2.0 mL11 × 40 mm10–22.00015.000Micro, Small, Medium, Large, FloorRS04, RS05no value
C-R-FA543636 × 0.5 mL (Micro 14, Medium 38) / 4 × PCR strip (Small 24)8 × 30 mm10–22.00015.000Micro, Small, Medium, Large, FloorNone (Micro/Medium) / RS04, RS05 (Small)Code-to-type matching is inconsistent between pages in the brochure; to be confirmed with the manufacturer
C-R-FA544848 × 0.2–0.4 mL6 × 40 mm10–22.00015.000Micro, Small, Medium, Large, Floorno valueno value
C-R-FA54944 × PCR strip (Micro 14, Medium 38) / 36 × 0.5 mL (Small 24)6 × 40 mm10–22.00015.000Micro, Small, Medium, Large, Floorno valueCode-to-type matching is inconsistent between pages in the brochure; to be confirmed with the manufacturer

Other products from the same brand